Titan Instructions for Use - EN
Reflex Functions
Signal sources
Tone - Contra, Reflex:
Tone - Ipsi, Reflex:
NB noise - Contra, Reflex:
NB noise - Ipsi, Reflex:
Noise - Contra, Reflex:
Noise - Ipsi, Reflex:
Outputs
Contra Earphone:
Ipsi Earphone:
Air:
Test types
Manual Reflex
Automated Reflex
Reflex Decay
Reflex latency
ABR Infant Screening
Preamplifier
One Channel:
Gain:
Frequency response:
Noise:
CMR Ratio:
Max input offset voltage:
Input impedance:
Power from main unit:
Electrical Impedance
Measurement frequency:
measurement
Waveform:
Measurement current:
Range:
Stimulus
Stimuli:
Stimulus rate:
Transducers:
(Calibrated to Standards)
Channels:
Level:
Bandwidth:
Recording
Analysis time:
A/D resolution:
Artifact reject system:
Display
Algorithmic Sensitivity
CE-Chirp®:
Specificity
CE-Chirp®:
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz.
500, 1000, 2000, 3000, 4000 Hz.
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz.
1000, 2000, 3000, 4000 Hz.
Wide Band, High Pass, Low Pass.
Wide Band, High Pass, Low Pass.
TDH39 earphone, DD45 earphone, CIR55 insert and/or EARtone
3A/IP30 insert for Reflex measurements.
Probe earphone incorporated in the probe system for Reflex
measurements.
Connection of the air system to the probe.
Manual control of all functions.
Automatic reflexes:
-
Single intensities
-
Reflex growth
Automatic, 10 dB above threshold and manually controlled with
stimulus durations of 10 to 30 s.
Automated, first 300 ms from stimulus start.
3 electrodes. 50 cm
Switchable: Software will automatically switch mastoid and ground if
mastoid montage is used. So the user does not need to replace
electrode during testing.
64 dB
0,5 - 5000 Hz
<25 nV/√Hz
>90 dB.
2.5 V
10 MΩ/ 170 pF
Isolated power supply
33 Hz
Rectangular
11.25 μA
0.5 kΩ – 25 kΩ ± 10 %
Click range (200 Hz -11 kHz)
CE-Chirp® range (200 Hz – 11 kHz)
HiLo CE-Chirp® range
(Lo – up to 1.5 kHz) & (Hi – above 1.5 kHz)
90 Hz
EARTone ABR insert phone
EARTone ABR for EarCup
IP30 ABR insert phone
IP30 ABR for EarCup
TDH 39 or DD45 head phone (Static force: 4,5N ± 0,5N
IOW Probe
2
30 dB nHL, 35 dB nHL, 40 dB nHL
22.05 kHz
1-10 min or Residual noise 5-80 nV
24 bit
Rejection level (Peak, Min RMS, Max RMS) & Clipping (Saturation)
Stimulus level and type, Bar and Graph view
Basic and advanced view
99.9%
> 96%
Page 92