Switching the Titan on and off

Switch the Titan on by pressing either the
To switch the Titan off, hold down both the
NOTICE
The Titan takes approximately 2 seconds to boot up. Allow the unit to warm up for 1 minute before use.

Probe Status

The probe status is indicated by the color of the light at the end of the extension cord, on the shoulder box or
on the probe status bar in the software. Below, the colors and their meanings are explained:
Color
Status
Red
Right ear is selected. Probe is out of the ear.
Blue
Left ear is selected. Probe is out of the ear.
Green
Probe is in the ear and a seal is maintained.
Yellow
Probe is in the ear and is blocked, leaking or too noisy.
White
The probe has just been attached. Probe status is unknown. If the probe light
stays white in any other situation the Titan may need to be switched off and
back on again to regain proper probe status.
Green changing
The current test has been completed.
to red/blue
No light
Titan is no longer monitoring the probe status, or the probe is in a big cavity
e.g when using Forced Start (for impedance only).

Titan Handheld Operation Panel

or
button.
and
button at the same time for 1 second.
Symbol
Function
F1 – F3
The F1 – F3 keys allow selection of various options
indicated on the Titan display screen above each
individual function key, e.g., "Print", "Save" and
"Delete".
The R and L buttons allow ear selection. Hold down
F4 – F5
F4 (R) and F5 (L) at the same time to turn off the
Titan.
F6 – F7
The Up and Down buttons (F6 and F7) navigate
through lists, menus etc. These buttons also allow
you to scroll through the different tests which are
indicated in the selected screen.