List of age groups will show when the protocol contains a wide band
tympanometry test. The selection in this list will by default match the
known age of the patient. Changing the age will impact which normative
data is displayed in absorbance graphs. Also, the calibration values of
wide band tympanometry tests depend on the age selection. Below and
above the age of six months the used calibration values change. Thus
make sure to have the correct age selected before the measurement
starts.
Show sketched absorbance examples allow to view examples of how
normal as well as pathologic absorbance curves could appear on
screen.
The hardware indication picture indicates whether the hardware is
connected. Simulation mode is indicated when operating the software
without hardware.
When opening the Suite the system will automatically search for the
hardware. If it does not detect the hardware a dialog box pops up and
asks if you want to continue in simulation mode.
The Timer symbol indicates when a measurement is running and when
it has stopped.
The protocol listing shows all tests that are part of the selected
protocol. The test that is displayed in the test screen area is highlighted
blue or red, depending on the chosen ear.
If more tests than can fit in the window are included in the protocol, a
scrollbar will be visible.
A checkmark in the box indicates that the test will run when START is
pressed. During testing the tests which have been completed will
automatically become unchecked. Uncheck the boxes of tests you do
not wish to run under the selected protocol before pressing START.
A white checkmark indicates that (at least some) data for this test is
stored in the memory.
START and STOP button are used for starting and stopping the
session.