Titan Instruction for Use - EN
1.3.1

Probe, Transducers and Cable Configurations

Titan operates with different transducers and cable configurations. The table below shows which transducers
and cable configurations are possible for use with the Titan.
Transducer/cable configuration
Probe directly connected to Titan
Probe + short extension cable
Probe + long clinical extension
cable (shoulder box)
Probe + PreAmplifier cable with
short extension cable
Probe directly connected to
PreAmplifier cable
1.3.2

The PreAmplifer Cable

The PreAmplifier cable (used with IMP440/DPOAE440/TEOAE440/ABRIS440) detects which ID transducer
is connected to it and will automatically read the calibration data from the ID transducer. The PreAmplifier
cable allows you to change from one ID transducer to another at any time as they use the special Omnetics
plug.
1.3.3

Contra Phone (for acoustic reflexes only)

The long clinical extension cable (shoulder box) used with IMP440/DPOAE440/TEOAE440 contains
calibration data for only one contra phone. The specific contra phone is calibrated with the shoulder box and
cannot be swapped over to another transducer without changing the calibration data in the shoulder box.
Available contra phones for use with the Long Clinical Extension Cable (shoulder box):
CIR55 (insert)
EARTone3A (single insert) (US only)
TDH39 (single headphone)
DD45 (single headphone)
IP30 contra with minijack
To run contralateral reflex measurements via the PreAmplifier cable, you will need a contra phone ID
transducer using the Omnetics plug. The contra phone that connects to the (IMP/OAE) shoulder box uses a
different plug and is not compatible with the PreAmplifier cable. Please contact your distributor if you require
a PreAmplifier cable compatible contralateral headset (ID transducer).
Available contra phones for use with the PreAmplifier cable:
CIR55 ID transducer (insert)
EarTone3A (single insert) ID transducer
TDH39 (single headphone) ID transducer
DD45 (single headphone) ID transducer
IP30 contra ID earphone
IMP440/WBT440
DPOAE440
Yes
Not permitted
Yes
Yes
Yes
Not permitted
Not permitted
TEOAE440
Not permitted
Yes
Yes
Yes
Yes
Yes
Yes
Not permitted
Page 3
ABRIS440
Not permitted
Not permitted
Not permitted
Yes