Titan Instructions for Use - EN
Patient noise displays the EEG peak values.
EEG displays the raw EEG.
Displays the EEG rejection level and the arrow buttons allow
modification of the rejection level
The electrode impedance values are displayed for the
corresponding electrode (white, red and blue).
Parameter display area shows the stimulus intensity and
stimulus type used for testing.
The infant picture displays the impedance status of the three
electrodes (good = green, amber = poor).
The transducer being used for testing is displayed (probe, insert
phone, EarCup or headphone).
The infant picture displays the impedance status of the three
electrodes (good = green, amber = poor). It also indicates the
montage for testing.
This picture indicates that the nape montage is required for
testing. This montage allows binaural testing (both ears at the
same time).
Connect the cables from the PreAmplifier as follows:
White cable: Vertex (hairline on the forehead)
Red cable:
Cheek
Blue cable:
Nape
This picture indicates that the mastoid montage is required for
testing.
Connect the cables from the PreAmplifier as follows:
White cable: Vertex (hairline on the forehead)
Red cable:
Right mastoid
Blue cable:
Left mastoid
Page 69