Table 1: Frequencies and Intensity Ranges for IMP440
Titan Maximums IMP
TDH39
Center
Reading
Freq.
Tone
NB
[Hz]
[dB HL]
[dB HL]
125
80
65
250
100
85
500
120
100
750
120
105
1000
120
105
1500
120
105
2000
120
105
3000
120
105
4000
120
105
6000
120
100
8000
105
95
10000
WB
-
120
LP
-
120
HP
-
120
Table 2: Frequencies and Intensity Ranges for DPOAE440
Titan Maximums DPOAE
IOW IPSI
IOW ch2
Reading
Reading
Center
Tone
Tone
Freq.
[Hz]
[dB SPL]
[dB SPL]
500
80
80
750
80
80
1000
80
80
1500
80
80
2000
80
80
3000
80
80
4000
80
80
6000
75
75
8000
65
65
10000
65
65
Titan Maximums TEOAE
Maximum TEOAE Click Intensity: 90 dB peSPL.
Titan Maximums ABRIS
Maximum ABRIS levels for Click and CE-Chirp® stimuli are limited to 30, 35 & 40 dBnHL for all transducers.
CIR55
EARtone 3A/IP30
Reading
Reading
Tone
NB
Tone
[dB HL]
[dB HL]
[dB HL]
85
70
100
100
85
110
110
100
115
110
105
120
115
105
120
115
105
120
115
105
120
115
105
120
110
100
120
95
95
105
80
85
90
-
120
-
-
120
-
-
120
-
IOW IPSI
Reading
NB
Tone
NB
[dB HL]
[dB HL]
[dB HL]
85
70
60
100
85
75
105
100
85
110
100
85
110
105
90
110
110
90
110
105
90
110
95
90
105
100
90
100
85
80
85
80
75
120
-
105
120
-
110
120
-
105
Page 95
DD45
Reading
NB
Tone
[dB HL]
[dB HL]
80
65
100
85
115
100
120
105
120
105
115
100
115
100
125
105
115
105
110
90
105
95
125
-
120
-
130
-

Advertisement