Page 75
Pointing at the desired graph and then using the scroll
wheel of your mouse allows zooming in and out on the
Response and DP-Gram graph. Once zoomed in on, the
graph then can be dragged in respect to the frequency axis.
Right clicking on the DP-Gram graph provides the following
options:
The scroll wheel of your mouse allows zooming in and out
with respect to the frequency axis. Additionally you can
Zoom in, Zoom out or Restore view by selecting the
appropriate item from the right mouse click menu.
Add an additional frequency after testing of the original
protocol has finished. Point and right click your mouse at the
frequency that you would like to test. Click on Add then
select an available frequency from the list to measure. After
adding one or more frequencies you will notice that the Start
button changes to say Continue. Clicking Continue will
measure all inserted frequencies without a time limit. Press
Stop when the additional points are sufficiently tested.
Next measurement point interrupts the automatic test
procedure and forces the Titan to start testing the next
frequency immediately. This function is available when max
test point has been selected in the protocol.
Measure only selected frequency results in re-testing the
currently selected measurement point only. Select a
measurement point to re-test by right clicking on it. A black
circle around the measurement point indicates that it is
selected. After pressing Continue (where the Start button
was) the selected point will be tested without a time
limitation. Press Stop to stop testing.
DPOAE norm data allows you to change which DP norm
data is shown in the DP-Gram.