Titan Instructions for Use - EN
Combined view or Single view button to toggle between combined
and single view. This icon is only available if during the protocol. Setup,
the Display wizard is used to make a Combined view.
List of Defined Protocols allows for selecting a test protocol for the
current test session (refer to the Additional Information document for
more details about protocols).
Temporary setup allows for making temporary changes to the selected
protocol. The changes will be valid for the current session only. After
making the changes and returning to the main screen, the name of the
protocol will be followed by an asterisk (*).
List of historical sessions accesses historical sessions for review or
the Current Session.
Go to current session takes you back to the current session.
Probe status is shown by a colored bar with description next to it.
When the probe status is Out of ear it will show the color of the selected
ear (blue for left and red for right). When the probe is detected to be In
ear the color is green. When Blocked, Leaking or Too Noisy the
colored bar is amber. When No probe is detected the status bar is
grey.
See section 3.3 for information about probe statuses.
Forced Start can be used to force and impedance measurement to
start when the probe status is not indicating 'in ear'. This can be used
for patients with PE tubes. Forced Start can be activated pressing the
icon or by long pressing the Start / Spacebar / button on shoulder
box for 3 seconds.
Either Auto or Manual testing can be selected. Manual testing means
that, where applicable the ear canal pressures can be set manually. In
or during reflexes you can also select individual stimuli separately or
add or delete stimulus types and levels. Although the option to test
manually normally is a global protocol setting, it does not influence the
way Eustachian tube function and wide band tympanometry tests are
performed. When turning back to automatic testing, you will be asked to
save data because some measurements might otherwise get lost due to
changes you may have made during testing.
Report editor button opens a separate window for adding and saving
notes to the current or historic session.
The pressure indicator shows how close the actual pressure is to
target pressure. This tool is only available when testing reflexes and
when testing wide band absorbance. The target pressure will in that
case in general be equivalent to the peak pressure of the tympanogram.
When the pressure drifts far away from the target pressure, the probe
status will change to Leaking.
Page 60