Interacoustics Titan Instructions For Use Manual page 43

Hide thumbs Also See for Titan:
Table of Contents

Advertisement

Titan Instructions for Use - EN
ABRIS440
DPOAE440
The circles on the baby's head indicate the electrode montage for the test
and the status of the electrode impedance; green = acceptable, amber =
poor.
The selected transducer name is displayed next to the baby's head.
Below the test time, the impedance value for each of the three electrodes
(white, red, blue) is displayed.
The three buttons on the bottom row of the screen correspond with the top
three function buttons on the hand held unit
During testing:
The EEG bar is displayed and shows the EEG peak values. The black bar
represents the EEG peak level above which a measurement will be rejected
(displayed in amber).
If the test is started when the electrode impedance indicators are amber
(poor), the impedance is not ideal warning window will appear. The user
must confirm whether they wish to proceed with testing if impedance values
are poor. Poor impedance can lead to longer test times and noisier
recordings.
The center of the screen is where the test
measurement is displayed.
The display will differ depending on the view
selected in the protocol: Bar view or Graph view.
Page 39