Wall Mounted Installation Of D1Cb (Electrical) - ProMinent DULCOMETER D1Cb Assembly And Operating Instructions Manual

Hide thumbs Also See for DULCOMETER D1Cb:
Table of Contents

Advertisement

1.
Using the dimensions shown in Fig. 4 mark the precise position of the device on
the control panel
2.
Mark the corner points and drill (drill diameter 12 - 13 mm)
3.
With a punching tool or jigsaw make the opening as per the punched template
drawing
4.
Chamfer the cut edges and check whether the sealing surfaces are smooth for the
caulk strip
ð Otherwise the seal cannot be guaranteed
5.
Press the caulk strip evenly into the groove running around the device
6.
Place the device into the control panel and fix in place at the rear by means of the
retaining brackets and PT cutting screws
ð The device should project approx. 35 mm from the control panel
3.4
Wall Mounted Installation of
D1Cb (Electrical)
WARNING!
Electrical voltage
Possible consequence: Fatal or very
serious injuries
The electrical connection to the
device should only be made once
it has been fitted to the wall or
control panel
The device must be electrically
disconnected before it is opened
Ensure that the device cannot be
reconnected accidentally
NOTICE!
Opening the device
Damage to the product or its sur‐
roundings
The device may only be opened
by qualified personnel
The device should only be
opened when fitted to the wall or
control panel
D1Cb mounting