Calibration Of The Sensor For Amperometric Measured Variables; Preparation For Calibration Of The Sensors For Amperometric Measured Variables - ProMinent DULCOMETER D1Cb Assembly And Operating Instructions Manual

Hide thumbs Also See for DULCOMETER D1Cb:
Table of Contents

Advertisement

Measured Variables and Operating Menus for Amperometric Sensors
Error message
Calibration not possible!
Zero point low
Calibration not possible!
Zero point high
* Please also note the operation manual for the respective sensor
7.6
Calibration of the Sensor for
Amperometric Measured
Variables
Only the slope can be calibrated in the
restricted operating menu of the
DULCOMETER
®
D1Cb / D1Cc.
Both the zero point and the slope can be
calibrated in the complete operating menu
of the DULCOMETER
Condition
< 3 mA
(only with 4 - 20 mA sen‐
sors)
> 5 mA
> 6 mA for 0.5 ppm chlorite
®
D1Cb / D1Cc.
Remarks *
Check sensor/cable
Repeat calibration in water
without metering medium
Check sensor/cable
Repeat calibration in water
without metering medium
7.6.1
Preparation for Calibration of
the Sensors for Ampero‐
metric Measured Variables
CAUTION!
Correct sensor operation / Run-in
time
Damage to the product or its sur‐
roundings
Correct measuring and dosing is
only possible if the sensor is
working perfectly
Please read the operating manual
for the sensor
Please also read the operating
manuals for the fittings and other
components used
It is imperative that the run in
periods of the sensors are
adhered to
The run in periods should be
allowed for when planning com‐
missioning
It may take a whole working day
to run-in the sensor
75