10.5
Set mA output
only with mA-Output
mA-Output
setting ?
Fig. 70: mA-Output
Set values of the mA-Output
Factory set‐
ting
Assign‐
Off
ment of
variable
Range
0 – 20 mA
Operating Menus Independent of Measured Variables
mA-Output
measured value
0...20 mA
error current
4...20 mA/21.5 mA
Possible values
Increment
Measured
value
Control
value
Correcting
value
Off
0-20 mA
4-20 mA
3.6/4-20 mA
control with mA-Output
control
control value:
positive ↑
negative ↓
mA-Output
measured value
0.00 ppm
0 mA =
20 mA =
1.00 ppm
mA-Output
correcting value
0 mA =
0.0 °C
20 mA =
100.0 °C
Lower value
Upper value Remarks
mA-Output
control value
0%
0 mA=
20 mA =
-100%
A0031
Possible if
control avail‐
able
Only available
with correc‐
tion variable
Reduction to
3.6 mA if
alarm relay is
switched
129