Measured Variables And Operating Menus For The Standard Signal General; Explanation Of The Standard Signal General - ProMinent DULCOMETER D1Cb Assembly And Operating Instructions Manual

Hide thumbs Also See for DULCOMETER D1Cb:
Table of Contents

Advertisement

Measured Variables and Operating Menus for the Standard Signal General

9
Measured Variables and Operating Menus for the Standard
Signal General
Users' qualification: instructed per‐
n
Ä Chapter 1.2 'Users' qual‐
sons, see
ifications' on page 10
9.1
Explanation of the Standard
Signal General
The measured variable "Standard Signal
General" of the DULCOMETER
D1Cc is used to connect sensors from
third party providers, which send a linear
mA signal, to the DULCOMETER
D1Cc. The DULCOMETER
is thus preconditioned to be able to
measure and control a number of physical
measured variables, using appropriate
Ä Table on page 105 .
sensors, see
®
D1Cb /
®
D1Cb /
®
D1Cb / D1Cc
Adjusting the sensor output signal
To match the DULCOMETER
D1Cc to the output signal from the sensor
or measuring device you must carry out
the following steps:
NOTICE!
Linear characteristic line
As the DULCOMETER
can only process linear characteristic
lines, only sensors/measuring devices
can be connected, which emit a linear
signal.
1.
Changing the DULCOMETER
D1Cb / D1Cc over to the complete
operating menu
2.
Set the required physical unit
(measured variable)
Display tolerances
With sensors or with output signals of
measuring devices, which do not
have to be calibrated or with which
calibration has to be done in the
sensor/measuring device, the display
tolerances must then be calibrated
between the sensor or measuring
device and the DULCOMETER
D1Cb / D1Cc.
3.
To do so, select the menu "Set
measured value", see
Ä Chapter 9.4 'Setting the
Reading ' on page 104 .
®
D1Cb /
D1Cb / D1Cc
®
®
®
101

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dulcometer d1cc

Table of Contents