Measured Variables and Operating Menus for the Standard Signal General
9.8
Calibrating the Standard Signal
In the restricted operating menu: The DULCOMETER
point.
In the complete operating menu: The DULCOMETER
point calibration.
WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries
–
The measuring range of the sensor or the measuring device is essential for the
measuring range!
–
If the assignment of the measuring range is modified, the settings must be
checked in all menus
–
If the assignment of the measuring range is changed, the sensor or the meas‐
uring device must be recalibrated
CAUTION!
Correct sensor operation / Run-in time
Damage to the product or its surroundings
–
Correct measuring and dosing is only possible if the sensor is working perfectly
–
Please read the operating manual for the sensor or the measuring device
–
It is imperative that the run in periods of the sensors or the measuring device are
adhered to
–
The run in periods should be allowed for when planning commissioning
–
It may take a whole working day to run-in the sensor or the measuring device
®
D1Cb / D1Cc calibrates the zero
®
D1Cb / D1Cc carries out a two-