Measured Variables and Operating Menus for the Standard Signal General
9.8

Calibrating the Standard Signal

In the restricted operating menu: The DULCOMETER
point.
In the complete operating menu: The DULCOMETER
point calibration.
WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries
The measuring range of the sensor or the measuring device is essential for the
measuring range!
If the assignment of the measuring range is modified, the settings must be
checked in all menus
If the assignment of the measuring range is changed, the sensor or the meas‐
uring device must be recalibrated
CAUTION!
Correct sensor operation / Run-in time
Damage to the product or its surroundings
Correct measuring and dosing is only possible if the sensor is working perfectly
Please read the operating manual for the sensor or the measuring device
It is imperative that the run in periods of the sensors or the measuring device are
adhered to
The run in periods should be allowed for when planning commissioning
It may take a whole working day to run-in the sensor or the measuring device
®
D1Cb / D1Cc calibrates the zero
®
D1Cb / D1Cc carries out a two-