Run 2 With Trial Mass In Plane 1 - Adash 4400 VA4 Pro User Manual

Vibration analyzer
Table of Contents

Advertisement

ADASH Ltd.
Adash 4400 – VA4Pro

RUN 2 with trial mass in plane 1

It is similar like single plane balancing, only two trial masses must be subsequently mount to two planes.
If you know the dynamic factor, then you can enter all 4 values manually and you do not need measure the
RUN2.
The red disc on the screen is displayed on the plane, to which should be just mount the trial mass.
If you defined the properties in Rotor Settings, then the recommended trial mass is displayed.
Use the Enter Trial button for re-entering of the trial mass, if it is required.
The corresponding plane for trial mass mounting is demonstrated by the red disc. After the measurement of
plane 1 take the measurement on plane 2 (or take them at once, when you use 2 channels). The trial mass
remain in the plane 1.
After measurements of both planes with trial mass on the plane 1 remove the trial mass from plane 1.
The RUN 2 with trial in plane 1 completed screen is in the next picture.