Band Fmin[Hz] - Hp Filtering; The Analyzer Buttons Description; The Arrow Mode Button - Adash 4400 VA4 Pro User Manual

Vibration analyzer
Table of Contents

Advertisement

ADASH Ltd.
Adash 4400 – VA4Pro
After the measurement is stopped (e.g. by Stop button), the computation continues until the input memory is
empty. The bellow window appears.
You can wait for all data acquisition or if you are not interested of measurement from input memory press the
Stop button to stop data processing.

Band fmin[Hz] - HP filtering

Almost in all measurements you should define the HP filtering of input signal. You set it in Band fmin[Hz]
parameter. You can select the none, 1, 2, 10 and user filter. The none value is available only for direct
measurement without integration. The none value does not mean the DC part measurement. The HP filter on
A/D converter is always used. But no additional filter is used. The frequency range with none filtering begins on
0.35 Hz (-3dB point). See the response graph.

The Analyzer buttons description

The Arrow Mode button

The Arrow Mode button switch functions of arrow buttons.
Try to push the Arrow Mode button and notice that the left top button changes between
Select/Position, Zoom X/Zoom Y and Move X/Move Y.
When the delta or sideband cursor is displayed, then the additional DeltaX item appears.
45
For further technical and contact information visit www.adash.eu, email: info@adash.cz