Bump Test
It enables to find the resonance frequency of machine body. Set the used channel and the number of hits for
averaging.
Test contains two parts. The level of amplitude trigger is found in the first part. The response spectrum is
developed in the second part.
Settings
Channel
Hits
Amplitude trigger setting
In the first part of test we need to detect the correct value for triggering. Make the hit and look the moving of the
hand on display. Make all hits with the same force. More then one hit is recommended (3-4). Push Stop to finish
it.