ADASH Ltd.
Adash 4400 – VA4Pro
Time signal measurement
Similar parameter like for overall is used. The red short vertical lines at the top of graph indicates the tacho
inputs.
For averaging you need to set the Global/ Trigger Settings/ Trigger Source = tacho. Then define the number of
averages.
You see the effect of averaging in decreasing of signal noise. The speed frequency is better seen. With the
cursor values are displayed also totRMS and tot0P values.
G-demod time signal measurement
This function enables to see the signal, when the standard envelope demodulation is used.
The Unit cannot be changed, only the acceleration g unit can be used. No integration of signal is enabled.
54
For further technical and contact information visit www.adash.eu, email: info@adash.cz