Adash 4400 – VA4Pro
ISO 10816 overall measurement
The ISO 10816 limit values are also available for overall measurements. The green/ amber-orange/ red point is
displayed in front of measured value according the ISO limit.
If you want to use this function you need to define the correct measurement parameters ( RMS velocity
measurement in 10-1000Hz range) and the group for machine ( see Global/ ISO Machine Settings).
The color of circle is related to the machine condition class A,B,C,D ( the green for A or B, the orange for C and
the red for D). Also the information about machine group and foundation are displayed. The limit which is
exceeding is also displayed at the bottom.
G-demod overall measurement
It is very similar to standard overall measurement. The signal is filtered in ( DEMOD fmin, DEMOD fmax) range
and demodulated then.
The Unit cannot be changed, only the acceleration g unit can be used. No integration of signal is enabled.
53