Step 2. Connect the load to the source output
Basic connections for sourcing voltage are shown in
t^okfkd
Step 3. Turn on voltage source
Press the OPER key to turn on the voltage source output. The VOLTAGE SOURCE
OPERATE indicator will turn on.
`^rqflk
A safety shield is advisable whenever measurements are being made
with voltages over 30V DC. Connections for the safety shield are
shown in
Figure
the noise shield or floating test circuit and it must be connected to
safety earth ground using #18 AWG or larger wire.
Do not connect external sources to the 6487 voltage source. External
sources may damage the 6487 voltage source.
Figure 3-4
Connections for sourcing voltage
V-SOURCE
OUTPUT
MADE IN
U.S.A.
CAT I
ANALOG OUT
DIGITAL I/O
!
505V PK
505V PK
505V PK
TRIGGER LINK
INPUT
RS-232
V-SOURCE OUTPUT
505V PK TO CHASSIS
!
LO
HI
INTERLOCK
FUSE
LINE
400mAT
100 VAC
505V
(SB)
120 VAC
MAX
220 VAC
200mAT
240 VAC
(SB)
Model 6487
Measurements and Sourcing Voltage
3-4. The metal safety shield must completely surround
Metal Safety Shield
Load
Safety
Earth
Ground
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
50, 60Hz
50 VA MAX
Figure
3-4.
3-17