Table 15-4
Reading limits for voltage source accuracy
Model 6487
Source Range
10V
50V
500V
Figure 15-3
Connections for voltage source output accuracy
Model 2001 DMM
DCV
ACV
DCI
ACI
PREV
DISPLAY
NEXT
REL
TRIG
STORE RECALL
POWER
INFO
LOCAL
CHAN SCAN
Verify voltage source accuracy for each of the values listed in
test point:
Select the correct Model 6487 voltage source range.
Set the voltage source output to the indicated value.
Make sure the voltage source is in operate (output on).
Verify that the DMM voltage reading is within the limits given in the table.
Repeat the procedure for negative source currents with the same magnitudes as
those listed in
Table
15-4.
Output
Voltage
10.000V
50.000V
500.00V
SENSE
INPUT
Ω 4 WIRE
HI
350V
1100V
!
PEAK
PEAK
2001 MULTIMETER
LO
500V
PEAK
INPUTS
Ω2
Ω4
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
FILTER MATH
2A 250V
AMPS
RANGE
CAL
CONFIG MENU
EXIT
ENTER
Model 6487 Output Voltage Limits
(1 Year, 18°C-28°C)
9.989 to 10.011V
49.946 to 50.054V
499.21 to 500.79V
INPUT
MADE IN
U.S.A.
CAT I
505V PK
505V PK
505V PK
INPUT
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
505V
MAX
V-SOURCE
OUTPUT
Performance Verification
Table
15-4. For each
Model 6487
IEEE-488
ANALOG OUT
(CHANGE IEEE ADDRESS
DIGITAL I/O
WITH FRONT PANEL MENU)
!
TRIGGER LINK
RS-232
!
HI INTERLOCK
FUSE
LINE
400mAT
100 VAC
(SB)
120 VAC
200mAT
220 VAC
(SB)
240 VAC
15-11
LINE RATING
50, 60Hz
50 VA MAX

Advertisement

Table of Contents
loading

Table of Contents