To allow many parallel connections to one instrument, stack the connector. Two screws are
located on each connector to ensure that connections remain secure. Current standards call
for metric threads, which are identified with dark-colored screws. Earlier versions had dif-
ferent screws, which were silver-colored. Do not use these types of connectors on the
Model 6487, because it is designed for metric threads.
Figure 9-2
To avoid possible mechanical damage, stack no more than three connectors on any one
unit.
klqb
shows a typical connecting scheme for a multi-unit test system.
Figure 9-2
Multi-unit connections
Instrument
To minimize interference caused by electromagnetic radiation, use only shielded
IEEE-488 cables. Available shielded cables from Keithley are Models 7007-1
and 7007-2.
Instrument
Instrument
Controller
Remote Operation
9-5