Keithley 6487 Reference Manual page 127

Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

Model 6487 Reference Manual
The 2-stage limit testing process is shown in . If Limit 1 fails, the L1 message is displayed
and the test is finished. Limit 2 is not tested because the pass band relationship between
the two stages implies that if Limit 1 fails, Limit 2 must also fail. If Limit 1 passes, the
Limit 2 test is performed. If Limit 2 fails, the L2 message is displayed. If both Limit 1 and
Limit 2 pass, the OK message is displayed. The display messages for limit tests are sum-
marized in
klqb
A test is only performed if it is enabled. Therefore, you can perform a single-stage test or a
2-stage test. In the flowchart (), operation simply proceeds through a disabled test.
Table
8-1.
When limit tests are enabled, the voltage source value display is not visible, but
it can still be viewed by using the CONFIG V-SOURCE menu or by pressing the
V-SOURCE up or down arrow keys.
Table 8-1
Test limit display messages
Display
Limit 1
Message
Test Result
:OK
Pass
:L1
Fail
:L2
Pass
Limit Tests and Digital I/O
Limit 2
Test Result
Pass
Not Performed
Fail
8-3