Keithley 6487 Reference Manual page 41

Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

Model 6487 Reference Manual
The 2V analog output signal is not corrected during calibration. Gain errors of up to 2.5%
may appear at this output, depending on range.
The output impedance is <100¾. To minimize the effects of loading, the input impedance
of the device connected to the ANALOG OUT should be as high as possible. For example,
for a device that has an input impedance of 1M¾, the error due to loading will be approxi-
mately 0.01%. High capacitance connected to the analog output will increase the rise time.
An internal 1k¾ resistance is connected between COM and analog common for protec-
tion. The effects of this resistance on analog output accuracy are negligible.
Rel and the result of mX+b, m/X+b, or LOG have no affect on the analog output. The 2V
analog output is scaled only to the actual input.
Figure 2-11
Typical analog output connections
A. Connections
MADE IN
U.S.A.
CAT I
505V PK
505V PK
505V PK
INPUT
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
505V
MAX
Model 6487 Rear Panel
B. Equivalent Circuit
Input from
Prescaler
IEEE-488
ANALOG OUT
(CHANGE IEEE ADDRESS
DIGITAL I/O
WITH FRONT PANEL MENU)
!
TRIGGER LINK
RS-232
!
HI
INTERLOCK
FUSE
LINE
400mAT
100 VAC
(SB)
120 VAC
200mAT
220 VAC
(SB)
240 VAC
A
Model 6487
Measurement Concepts and Connections
LINE RATING
50, 60Hz
50 VA MAX
Test Lead
<100Ω
Analog Output
COM
1kΩ∗
A
Measuring Device
(i.e. Chart recorder)
LO
HI
* 1kΩ included for protection.
Virtually 0Ω for accuracy purposes.
R
= Input Resistance of
L
measuring device
2-15