Keithley 6487 Reference Manual page 204

Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

Model 6487 Reference Manual
D) SYSTem:VERSion
Read the version of the SCPI standard being used by Model 6487. Example response
message: 1996.0.
E) SYSTem:KEY <NRf>
Parameters
This command is used to simulate front panel key presses. For example, send the follow-
ing command to simulate pressing the "MATH" key: SYSTem:KEY 3. The queue for the
:SYST:KEY? query command can only hold one key-press and this query cannot be used
to determine keys physically pressed from the front panel.
When :SYST:KEY? is sent and Model 6487 is addressed to talk, the key-press code num-
ber for the last :SYST:KEY command is sent to the computer. The value is -1 if a
:SYST:KEY command has not been sent since the last time the unit was placed in remote.
Note that sending :SYST:KEY 1 will simulate a press of the CONFIG key, not LOCAL.
To place the instrument in the local state, send the GTL command as described in
Section
1 = CONFIG/LOCAL key
2 = I | Ω key
3 = MATH key
4 = FILT key
5 = ZCHK key
6 = REL key
7 = OPER key
8 = SOURCE up arrow key
9 = -------
10 = -------
11 = RANGE up arrow key
12 = AUTO key
13 = RANGE down arrow key
14 = ENTER key
15 = Cursor right arrow key
16 = SOURCE down arrow key
9.
DISPlay, FORMat, and SYSTem
17 = MENU key
18 = COMM key
19 = DISP key
20 = TRIG key
21 = LIMIT key
22 = DIGITS key
23 = RATE key
24 = Cursor left arrow key
25 = -------
26 = SAVE key
27 = SETUP key
28 = STORE key
29 = RECALL key
30 = AZERO key
31 = DAMP key
32 = EXIT key
13-11

Advertisement

Table of Contents
loading