Keithley 6487 Reference Manual page 178

Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

10-14
Status Structure
Questionable event status
The used bits of the questionable event register
Bit B3, low limit 2 fail (LL2F) — Set bit indicates that the low limit 2 test has
failed.
Bit B4, high limit 2 fail (HL2F) — Set bit indicates that the high limit 2 test has
failed.
Bit B5, limits pass (LP) — Set bit indicates that all limit tests passed.
Bit B6, reading available (RAV) — Set bit indicates that a reading was taken and
processed.
Bit B7, reading overflow (ROF) — Set bit indicates that the reading exceeds the
selected measurement range of the Model 6487.
Bit B8, buffer available (BAV) — Set bit indicates that there are at least two
readings in the buffer.
Bit B9, buffer full (BFL) — Set bit indicates that the buffer is full. This bit will
also be set when a voltage sweep has been completed
programmed number of A-V ohms cycles have been taken
Bit B10, input overvoltage (IOV) — Set bit indicates there is an input over
voltage condition.
Bit B11, output interlock asserted (INT) — Set bit indicates that the output
interlock is asserted and the voltage source output cannot be turned on.
Bit B14, voltage source compliance (VSC) — Set bit indicates that the voltage
source is in compliance.
Bit B7, calibration summary (Cal) — Set bit indicates that an invalid calibration
constant was detected during the power-up sequence. This error will clear after
successful calibration of the Model 6487.
Bit B14, command warning (Warn) — Set bit indicates that a signal oriented
measurement command parameter has been ignored.
Model 6487 Reference Manual
(Section
(Section
(Figure
10-7) are described as follows:
6), and if the
3).