C-6
DDC Emulation Commands
Table C-1 (cont.)
Device dependent command summary
Mode
Command
N0
Data Store
Nn
O0
Operate
O1
P0
P1
Filters
P2
P3
Q0
Interval
Q1
R0
R1
R2
R3
R4
R5
Range
R6
R7
R8
R9
R10
Equivalent SCPI or 488.2 commands
Stop storage, clear buffer
Arm data store, set buffer size "n" where n =1 to 512
Place voltage source in standby
SOURce[1]:VOLTage:STATe OFF
Place voltage source in operate
SOURce[1]:VOLTage:STATe ON
Both analog and digital filters off
Enable digital filter, disable analog filter
Disable digital filter, enable analog filter
Enable both digital and analog filters
175msec
TRIGger:DELay 0.175
Set to "n" seconds, n= 0.01msec to 999.999sec
TRIGger:DELay <NRf>
Enable autorange
[SENSe[1]]:RANGe:AUTO ON
2nA range
[SENSe[1]]:RANGe 2e-9
20nA range
[SENSe[1]]:RANGe 20e-9
200nA range
[SENSe[1]]:RANGe 200e-9
2μA range
[SENSe[1]]:RANGe 2e-6
20μA range
[SENSe[1]]:RANGe 20e-6
200μA range
[SENSe[1]:RANGe 200e-6
2mA range
[SENSe[1]]:RANGe 2e-3
20mA range
[SENSe[1]]:RANGe 20e-3
20mA range
[SENSe[1]]:RANGe 20e-3
Disable autorange
[SENSe[1]]:RANGe:AUTO OFF
Description
Model 6487 Reference Manual
Note
Default
J
J
K
K
L
M
M
M
X
X
X
X

Advertisement