Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

Model 6487 Reference Manual
As shown in
capacitance of the picoammeter (A). Picoammeter HI is connected to the higher resistance
side of the circuit being measured, the "R" side of this current source. This circuit is con-
nected properly, so this current does not flow through the picoammeter and, therefore,
does not cause any measurement errors.
However, when the HI of the picoammeter is connected to the low impedance side of the
DUT, this AC current (i) flows through the picoammeter (A) as illustrated in
This current may affect the measurement accuracy, especially at low signal levels.
Figure
G-8, this will cause a current (i) to flow through the low to ground
Figure G-8
Proper connection
Current Source
R
+
-
i
Applications Guide
HI
6487
A
Picoammeter
LO
G-11
Figure
G-9.