6487 Picoammeter Specifications
5
DIGIT
ACCURACY (1YR)
DEFAULT
±(% RDG. + OFFSET)
RANGE RESOLUTION 18°–28°C, 0–70% RH
2 nA
10 fA
0.3 % + 400 fA
20 nA
100 fA
0.2 % +
200 nA
1 pA
0.15 % + 10 pA
2 µA
10 pA
0.15% + 100 pA
20 µA
100 pA
0.1 % +
200 µA
1 nA
0.1 % + 10 nA
2 mA
10 nA
0.1 % + 100 nA
20 mA
100 nA
0.1 % +
TEMPERATURE COEFFICIENT: 0°–18°C & 28°–50°C. For each °C, add 0.1 × (% rdg + offset) to
accuracy spec.
INPUT VOLTAGE BURDEN: <200µV on all ranges except <1mV on 20mA range.
MAXIMUM INPUT CAPACITANCE: Stable to 10nF on all nA ranges and 2µA range; 1µF on 20µA
and 200µA ranges, and on mA ranges.
MAXIMUM CONTINUOUS INPUT VOLTAGE: 505 VDC
NMRR
1
: (50 or 60Hz) :60dB
ISOLATION (Ammeter Common or Voltage Source to chassis): Typically >1×10
with <1nF.
MAXIMUM COMMON MODE VOLTAGE (Between Chassis and Voltage Source or Ammeter):
505 VDC.
MAXIMUM VOLTAGE BETWEEN VOLTAGE SOURCE AND AMMETER: 505 VDC
ANALOG OUTPUT: Scaled voltage output (inverting 2V full scale on all ranges) 2.5% ±2mV
ANALOG OUTPUT IMPEDANCE
3
: <100Ω, DC-2kHz.
VOLTAGE SOURCE
Range
Step Size
Accuracy
5
(Max)
(typical)
±(% PROG. + OFFSET)
18°C - 28°C, 0 - 70% R.H.
±10.100
200µV
0.1% + 1mV
±50.500
1mV
0.1% + 4mV
±505.00
10mV
0.15% + 40mV
SELECTABLE CURRENT LIMIT: 2.5mA, 250µA, 25µA for 50V and 500V ranges, 25mA additional
limit for 10V range. All current limits are -20%/+35% of nominal.
WIDEBAND NOISE
9
: <30mVp-p 0.1Hz - 20MHz.
TYPICAL TIME STABILITY: ±(0.003% + 1mV) over 24 hours at constant temperature (within 1°C,
between 18°C - 28°C, after 5 minute settling).
OUTPUT RESISTANCE: <2.5Ω.
VOLTAGE SWEEPS: Supports linear voltage sweeps on fixed source range, one current or resist-
ance measurement per step. Maximum sweep rate: 200 steps per second. Maximum step count
3000. Optional delay between step and measure.
RESISTANCE MEASUREMENT (V/I): Used with voltage source; resistance calculated from voltage
setting and measured current. Accuracy is based on voltage source accuracy plus ammeter
accuracy. Typical accuracy better than 0.6% for readings between 1kΩ and 1TΩ.
ALTERNATING VOLTAGE RESISTANCE MEASUREMENT: Offers alternating voltage resistance
measurements for resistances from 10
voltage up to ±505V.
1
At 1 PLC – limited to 60 rdgs/sec under this condition.
2
At 6 PLC, 1 standard deviation, 100 readings, filter off, capped input – limited to 10 rdgs/sec
under this condition.
3
Measured at analog output with resistive load >2kΩ.
4
Maximum rise time can be up to 25% greater.
5
Accuracy does not include output resistance/load regulation.
6
Rise Time is from 0V to ± full-scale voltage (increasing magnitude).
7
Fall Time is from ± full-scale voltage to 0V (decreasing magnitude).
8
For capacitive loads, add C*ΔV/I
9
Measured with LO connected to chassis ground.
1
TYPICAL
RMS NOISE
2
20 fA
1 pA
20 fA
1 pA
1 pA
1 nA
100 pA
100 pA
10 nA
1 µA
10 nA
Noise (p-p)
Temperature
0.1 - 10 Hz
Coefficient
<50µV
(0.005% + 20µV) / ˚C
<150µV
(0.005% + 200µV) / ˚C
<1.5mV
(0.008% + 2mV) / ˚C
Ω to 10
Ω. Alternates between 0V and user-selectable
9
15
to Rise Time, and C*ΔV/1mA to Fall Time.
Limit
TYPICAL ANALOG
RISE TIME (10% TO 90%)
3
IEEE-488 BUS IMPLEMENTATION: SCPI (IEEE-488.2, SCPI-1996.0);
DAMPING
4
DDC (IEEE-488.1).
OFF
ON
LANGUAGE EMULATION: Keithley Model 486/487 emulation via
4 ms
80 ms
DDC mode.
4 ms
80 ms
RS-232 IMPLEMENTATION:
300 µs
1 ms
Supports: SCPI 1996.0.
300 µs
1 ms
Baud Rates: 300, 600, 1200, 2400, 4800, 9600, 19.2k, 38.4k, 57.6k.
110 µs
110 µs
Protocols: Xon/Xoff, 7 or 8 bit ASCII, parity-odd/even/none.
110 µs
110 µs
Connector: DB-9 TXD/RXD/GND.
110 µs
110 µs
110 µs
110 µs
AMMETER INPUT CONNECTOR: Three lug triaxial on rear panel.
ANALOG OUTPUT CONNECTOR: Two banana jacks on rear panel.
VOLTAGE SOURCE OUTPUT CONNECTOR: Two banana jacks on rear
panel.
INTERLOCK CONNECTOR: 4 pin DIN.
TRIGGER LINE: Available, see manual for usage.
DISPLAY: 12 character vacuum fluorescent.
Ω in parallel
11
DIGITAL FILTER: Median and averaging (selectable from 2 to 100
readings).
RANGING: Automatic or manual.
AUTORANGING TIME
OVERRANGE INDICATION: Display reads "OVRFLOW".
CONVERSION TIME: Selectable 0.01 PLC to 60 PLC (50PLC under
50Hz operation). (Adjustable from 200µs to 1s)
READING RATE:
To internal buffer 1000 readings/second
Typical Rise
Typical Fall
Time
6,8
Time
7,8
To IEEE-488 bus
(10%-90%)
(90%-10%)
BUFFER: Stores up to 3000 readings.
250 µs
150 µs
PROGRAMS: Provide front panel access to IEEE address, choice of
250 µs
300 µs
engineering units or scientific notation, and digital calibration.
4.5 ms
1 ms
EMC: Conforms with European Union Directive 89/336/EEC,
EN61326-1.
SAFETY: Conforms with European Union Directive 73/23/EEC,
EN61010-1, CAT I.
ENVIRONMENT:
Operating: 0°–50°C; relative humidity 70% non-condensing, up to
35°C. Above 35°C, derate humidity by 3% for each °C.
Storage: –10°C to +65°C.
WARM-UP: 1 hour to rated accuracy (see manual for recommended
procedure).
POWER: 100–120V or 220–240V, 50–60Hz, (50VA).
PHYSICAL:
Case Dimensions: 90mm high × 214mm wide × 369mm deep (3
Working Dimensions: From front of case to rear including power
NET WEIGHT: <4.7 kg (<10.3 lbs).
Notes:
1
0.01 PLC, digital filters off, front panel off, auto zero off.
2
Binary transfer mode.
3
Measured from trigger in to meter complete
Specifications are subject to change without notice.
REMOTE OPERATION
GENERAL
3
: <250ms (analog filter off, 1PLC)
900 readings/second
× 8
in. × 14
3
9
in.).
8
16
cord and IEEE-488 connector: 394mm (15.5 inches).
IEEE-488.1.
1
1, 2
1
in.
2
.
Rev. A
HW 10/25/02

Advertisement

Table of Contents
loading

Table of Contents