Keithley 6487 Reference Manual page 305

Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

F-4
Remote Calibration
Table F-2
Current calibration commands by range
Range
2nA
:SENS:CURR:RANG 2e-9
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-9
:CAL:PROT:SENS –2e-9
20nA
:SENS:CURR:RANG 2e-8
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-8
:CAL:PROT:SENS –2e-8
200nA
:SENS:CURR:RANG 2e-7
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-7
:CAL:PROT:SENS –2e-7
2μA
:SENS:CURR:RANG 2e-6
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-6
:CAL:PROT:SENS –2e-6
20μA
:SENS:CURR:RANG 2e-5
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-5
:CAL:PROT:SENS –2e-5
200μA
:SENS:CURR:RANG 2e-4
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-4
:CAL:PROT:SENS –2e-4
2mA
:SENS:CURR:RANG 2e-3
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-3
:CAL:PROT:SENS –2e-3
20mA
:SENS:CURR:RANG 2e-2
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2e-2
:CAL:PROT:SENS –2e-2
* Full-scale calibration values for 2nA to 2μA ranges should be calcu-
lated from calibrator voltages and standard resistance values. (See
Section
16.) Values for 20μA to 20mA ranges can be used as shown.
Model 6487 Reference Manual
Commands*