Programming Example - Keithley 6487 Reference Manual

Picoammeter/voltage source
Hide thumbs Also See for 6487:
Table of Contents

Advertisement

Model 6487 Reference Manual
If the instrument is programmed to perform an infinite number of measurements (arm
count or trigger count set to infinite), you cannot use the :DATA? command to return
CALC2 readings. However, you can use the :DATA:LATest? command to return the last
CALC2 reading after aborting the measurement process. After sending the INITiate com-
mand to start the measurement process, use the ABORt command to abort the measure-
ment process, then use :DATA:LATest? to return the last CALC2 reading.
Sending :DATA? or :DATA:LATest? without first sending INITiate will return "old" read-
ings or cause an error (-220) if limit is not enabled or there are no readings available.
E) :ARM:SOURce <name>
Typical "start of test" options:

Programming example

The following command sequence will test DUT using the limit tests example shown in
Figure
*RST
CALC2:LIM:UPP 2e-3
CALC2:LIM:LOW -2e-3
CALC2:LIM:STAT ON
CALC2:LIM2:UPP 1e-3
CALC2:LIM2:LOW -1e-3
CALC2:LIM2:STAT ON
SYST:ZCH OFF
INIT
CALC2:LIM:FAIL?
CALC2:LIM2:FAIL?
IMMediate — Test starts when LIMIT key is pressed.
NSTest — Test starts when component handler pulls the SOT line low.
PSTest — Test starts when component handler pulls the SOT line high.
BSTest — Test starts when component handler pulls the SOT line high or low.
8-2.
Limit Tests and Digital I/O
' Restore RST defaults.
' Set upper limit for Limit 1 (2mA).
' Set lower limit for Limit 1 (-2mA).
' Enable Limit 1 test.
' Set upper limit for Limit 2 (1mA).
' Set lower limit for Limit 2 (-1mA).
' Enable Limit 2 test.
' Connect DUT to input.
' Disable zero check.
' Perform tests on DUT (one measurement).
' Return result of Limit 1 test.
' Return result of Limit 2 test.
8-21