Loading Base Image Software
The second choice is to load only the bootable portion of the hard drive partition, the "C" partition. This
is not an option for a new hard drive that has not been previously partitioned for proper LOGIQ 9
operation.
2.) Select [2} Load the bootable C: partition only - Patient data is NOT lost.
NOTE:
DO NOT USE THIS OPTION ON A BRAND NEW SYSTEM. IT IS INTENDED FOR
RECOVERY OF A SYSTEM THAT WILL NOT BOOT UP. APPLICATION SOFTWARE WILL
NEED TO BE INSTALLED WHEN THIS PROCEDURE HAS COMPLETED. ALL PATIENT
DATA IS PRESERVED.
This selection will leave the remaining drive partitions intact.
Table 8-7
Drive Partition
Figure 8-68
or abort by pressing CTRL-C.
The third choice,
3.) Exit to the A:\ prompt. Is intended for manufacturing use only.
This is NOT an option as it is intended for manufacturing use only.
If you select Option 3 by mistake, power cycle the system. Hold down the Standby switch until the light
turns amber then turn the system back on.
8-62
PROPRIETARY TO GE
3
EVISION
LOGIQ 9 Drive Partition and Usage
C
System Files and Applications
D
User Define and Presets
E
Patient Archive Data
F
File Swap Area
Figure 8-68 Base Load Option 2 Message
is the last warning message before the base load process begins. Press any key to continue
Section 8-12 - Loading Base Image Software
(cont'd)
Usage