GE M
S
EDICAL
YSTEMS
D
2294854-100, R
IRECTION
8-16-6
If the System is NOT on a Network
If the system is connected to a network, skip this section.
If the system is NOT on a network, a static "dummy" IP-Address must be install in Windows for DICOM
and Daignostics to function.
To install a Static IP-Address in windows:
1.) Boot up the system with the service dongle attached to the USB port.
2.) From the Start Applications dialog box, like the one shown in
3.) The Maintenance Password screen will appear. Enter the current Service Password.
4.) Select Exit to Windows.
5.) From the task bar at the bottom of the Touch Panel select START>SETTINGS>NETWORK
6.) From the window shown in
8-86
PROPRIETARY TO GE
3
EVISION
MAINTENANCE Mode to enter the Windows desktop.
Figure 8-95 Select Maintenace Mode
DIALUP CONNECTIONS.
Figure 8-96 Network and Dialup Connections Window
Section 8-16 - Optional Software Procedures
Figure
8-96, double click on LOCAL AREA CONNECTIONS.
LOGIQ™ 9 P
ROPRIETARY
Figure
8-95, select
M
ANUAL