You can select one of the following labels for each temperature channel:
Source Label
ART
CORE
ESOPH
RECT
SKIN
VEN
NASO
If you don't select a label, the Temperature channels appear with the default labels of T1 and
T2.

Temperature System Messages

The Propaq M unit may display the following messages when monitoring Temperature.
Note:
System Message
CHECK PROBE
PROBE FAILURE
TEMP DISABLED
9650-000820-01 Rev. K
Description
Arterial temperature probe
Central body temperature or tympanic membrane temperature
probe
Esophageal temperature probe
Rectal temperature probe
Skin temperature probe (surface application)
Ventilator airway temperature probe
Nasopharynx or nasal/oral temperature probe
The temperature function performs a self test when initially powered on and also
performs system tests automatically, every 10 seconds, while this function is active.
Propaq M Operator's Guide
Temperature System Messages
Cause
The temperature probe is disconnected. Check probe
and reconnect it.
The temperature probe is defective. Replace the
Temperature probe.
A system error has occurred. The Propaq M unit cannot
take temperature measurements and should be serviced.
12-5