Recommended And Alternate Ultrasound Room Layout - GE LOGIQ F Series Basic Service Manual

Hide thumbs Also See for LOGIQ F Series:
Table of Contents

Advertisement

GE
D
5446617-100, R
IRECTION
2-3-3

Recommended and Alternate Ultrasound Room Layout

Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
(76 CM)
2 - 8
10
EVISION
FILM VIEWER
COUNTER TOP
SECRETARYS OR
DOCTOR'S DESK
FILM VIEWER
LINEN SUPPLY
SINK
PROBES/SUPPLIES
FILM SUPPLIES
EXTERNAL
PERIPHERALS
FOOT
LOGIQ 5
SW
CONSOLE
STOOL
EXAMINARION
TABLE
24 IN.
76 IN.
(61 CM)
(193 CM)
DOOR
30 IN.
An 8 by 10 foot Minimal Floor Plan
Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT
Section 2-3 - Facility Needs
DEDICATED
POWER
RECEPTACLE
SINK
SUCTION LINE
EMERGECY
OXYGEN
OVERHEAD
LIGHTS DIMMER
DOOR
42 IN.
(107 CM)
A 14 by 17 foot Recommended Floor Plan
Scale : Each square equals one square foot
DEDICATED POWER
OUTLETS
DEDICATED ANALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
GE CABINET
FOT SOFTWARE
AND MANUALS
B
ASIC
DEDICATED ALALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
18 IN.
LINEN SUPPLY
(46 CM)
CONSOLE
FOOT
SW
STOOL
76 IN.
(193 CM)
24 IN.
(61 CM)
PATIENT
TOILET
FACILITY
LOGIQ F S
ERIES
S
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents