I/O Signal Circuit Diagrams
NPN
Brown Power supply (24 VDC)
Black
Orange
Light blue
Blue
Pink
Gray
Green
Red
White
Purple
Yellow
Important
Preventing Chattering
• The Sensor is equipped with an anti-chattering function, but if the chattering is 100 s or longer, a faulty input may
occur. (Input signals of 99 s or shorter are ignored. Signals of 100 s or longer are treated as input signals.)
• Use no-contact output devices (e.g., SSR or PLC transistor output) for the input signals. If contacts (e.g., relay) are
used, chattering may cause the trigger to be input again during execution of a measurement.
Power Supply Specifications When a Switching Regulator Is Connected
Use a power supply that meets the following specifications. (The power supply is sold separately.)
Item
Power supply voltage
Output current
Recommended Power Supply
External power supply terminal screws
Important
Supply power from a DC power supply for which measures have been applied to prevent high voltages (e.g., a safety
extra low voltage circuit).
If UL certification is required for the overall system, use a UL Class II DC power supply.
Attaching the LED Warning Label
Attach the enclosed LED warning label to the cable or other location. The LED warning label must be attached
to a location that is readily visible from the Sensor.
Attachment Example
Warning Label
10-5 LED Safety p. 188
FQ User's Manual
Load
OUT0 (OR)
OUT1 (BUSY)
OUT2 (ERROR)
24 VDC
GND (0V)
TRIG
IN0
IN1
IN2
IN3
IN4
IN5
Description
24 VDC (21.6 to 26.4 V)
3.75 A min.
S8VS-09024@ (24 VDC, 3.75 A)
M4 (tightening torque: 1.2 N·m)
Warning Label
PNP
Brown Power supply (24 VDC)
Pink
Gray
Green
Red
White
Purple
Yellow
Light blue
Orange
Black
Blue
TRIG
IN0
IN1
IN2
IN3
IN4
IN5
24 VDC
OUT2 (ERROR)
OUT1 (BUSY)
OUT0 (OR)
Load
GND (0V)
27
Wiring

Advertisement

Table of Contents
loading

Table of Contents