Note
The field separator is not output unless the data continues.
The following range of values can be output.
999,999,999.9999 Measured value 999,999,999.9999
If the measured value is lower than 999,999,999.9999, then 999,999,999.9999 is output.
If the measured value is higher than 999,999,999.9999, then 999,999,999.9999 is output.
The following values are output if JG (Judge) is set.
OK: 0
NG: 1
Note
Data that is output after measurement is output until the last data even after the measurement is finished. Data out-
put is not interrupted midway.
When Output Format Is Binary
Set the numerical expression.
Select either fixed decimal or floating-point decimal.
Output Format
<Measured value of data 0 × 1,000><Measured value of data 1 × 1,000>
4 bytes
The measurement data multiplied by 1,000 is output continuously at 4 bytes per data. Negative numbers are
output as two's complements.
Example: When Data 0 Is 256.324 and Data 1 Is 1.000.
$ 0 0
$ 0 3
$ E 9
$ 4 4
Data 0: 256324
(256.324 × 1000)
Note
Binary output does not use data separators, i.e., field separators or record separators. These separators are used
only for ASCII output.
The following range of values can be output.
2,147,483.648 Measured value 2,147,483.647
If the measured value is lower than 2,147,483.648, then 2,147,483.648 is output.
If the measured value is higher than 2,147,483.648, then 2,147,483.648 is output.
The following values are output if JG (Judge) is set.
0 ( 0 1000)
OK:
NG: 1000 (1 1000)
Note
Data that is output after measurement is output until the last data even after the measurement is finished. Data out-
put is not interrupted midway.
FQ User's Manual
4 bytes
$ F F
$ F F
$ F C
$ 1 8
−
Data 1:
1000
A delimiter will
(−1.000 × 1000)
be attached to
the end.
<Measured value of data 7 × 1,000>
· · ·
4 bytes
$ 0 D
Outputting/Controlling with Ethernet
CR
Delimiter