4 - 10 MP6000 INTEGRATOR GUIDE
Verification Tests for Metric Scale (Continued)
Table 4-5
Verification Test for
Metric Scale Applied
Load kg
5.000
2.500
0.100
Return to Zero Test
0.000
Shift Test (see
Figure
5.00
Position 1
5.00
Position 2
5.00
Position 3
5.00
Position. 4
The weight checked at each position on the scale, shown in
of the platter and corner. Use a stack configuration of weights if needed (i.e., (2) 5.00 lb weights, or (2) 2.500 kg
weights) depending on the units of measure programmed.
Weight Positions on the Scale Platter
Figure 4-4
Indication = Applied
Load Within
Applicable
Tolerances
5.000
2.500
0.100
0.000
4-4)
5.000
5.000
5.000
5.000
Shaded Area
Represents the
Horizontal Window
Scale Platter
All Tolerances
All Tolerances
Are +/-
Are +/-
Acceptance
Acceptance
Single Int
Dual Int
.005 kg
.003 kg
.0025 kg
.002 kg
.0025 kg
.001 kg
0.0 kg
0.0 kg
.005 kg
.003 kg
.005 kg
.003 kg
.005 kg
.003 kg
.005 kg
.003 kg