4 - 4
MP6000 INTEGRATOR GUIDE

Legal Scale Dampening Filter

Set the scale sensitivity to vibration by scanning the appropriate bar code (see page 4-17). To program this
parameter, the scale must be in Calibration Mode. Bar code command values are as follows:
0 = Higher (most sensitive to vibration)
1 = Low (factory default)
2 = Very Low
3 = Ultra Low
The lower the number, the more sensitive the scale is to vibration. The higher the number, the slower the scale
may operate.

Step 3 - Calibration at NO LOAD

When Calibration Mode enters properly, the Scale Display blinks between CAL00 and CAL _ _ and the
7-segment display scrolls
Ensure the platter is installed and there is NO WEIGHTED LOAD on it.
1.
Touch the Scale Zero button on the front panel.
2.
If Calibration at NO WEIGHTED LOAD is successful after the Scale Zero button is pressed, the process
3.
continues with
IMPORTANT The steps in
or
If Calibration at NO WEIGHTED LOAD is not successful after the Scale Zero button is pressed, the
process continues with

Step 4 - Calibration at LOAD

IMPORTANT These steps must be performed within 90 seconds after pressing the Scale Zero button in
The Scale Display blinks either CAL25 or CAL11, depending on the units of measure programmed
(CAL25 = pounds; CAL11 = kilograms).
The diagnostic 7 -segment display scrolls C25Lb or C11g.
To continue with calibration:
C00Lb,
C00g
or
. Scale calibration can begin.
Step 4 - Calibration at LOAD on page
Step 4 - Calibration at LOAD
Scale Zero button.
Calibration Failure on page
Calibration at NO
LOAD.
4-4.
must be performed within 90 seconds after pressing the
4-7.
Step 3 -