Stream Mode; Preparation And Performance (Examples) - A&D HW-10KGL Instruction Manual

Hv-g/hw-g series digital platform scale
Hide thumbs Also See for HW-10KGL:
Table of Contents

Advertisement

16.2. Stream Mode

The scale outputs the current weighing data at the time the display is refreshed.
The scale does not output data while in the setting mode.
Averaging of function table
f13 0 Fast response
Sensitive value
f13 1
ff13 2 Slow response
Stable value
16.2.1. Preparation and Performance
(Examples)
Step 1 Turn off the display.
Press and hold the
key. The function table is displayed.
Step 2 Press the
Select a parameter of "Baud rate" with the
Step 3 Press the
mode").
Select "Stream mode ( f5 0 )" with the
Step 4 Press the
Select "Format 1 ( f15 0 )" with the
Step 5 Press the
Step 6 Press the
The scale returns to the normal weighing mode and
outputs the data continuously.
HV-G/HW-G Series
Approximately 7 times/s while the display is unstable,
Approximately 4 times/s when the display is stable
Approximately 4 times/s
key and press the
ZERO
key to display f4 ("Baud rate").
ENTER
key to display
ENTER
key to display f15 ("Format").
ENTER
key to display end .
ENTER
key to store the new parameters.
ENTER
Refresh rate
ON/OFF
key.
f5
("Output
key.
key.
Page 59