14.2. Complete Calibration Procedure

14.2.1. Gravity Acceleration Correction

Step 1 Turn on the display.
Open the rear cover of the display unit. Locate the
CAL switch inside.
Press and hold the
mode.
Then
Cal 0
Step 2 Press the
correction mode.
Step 3 Set your local gravity acceleration using the following
keys.
key
key
Step 4 Press the
Then
Cal 0
Step 5 Press the
the normal weighing mode.

14.2.2. Preparation

Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and stable power.
Install the scale on a solid floor where there is no draft, vibration, strong magetic
fields or direct sunlight.
Refer to "4.1. Installing the Scale".
Step 7 Keep the display turned on for at least 30 minutes to warm up the scale.
HV-G/HW-G Series
switch to enter the calibration
CAL
is displayed.
key to enter the gravity acceleration
ZERO
Selecting the figure to be changed.
Changing the number of the figure.
key to store the new value.
ENTER
is displayed.
switch again. Then the scale returns to
CAL