Calibration (Adjusting The Scale) - A&D HW-10KGL Instruction Manual

Hv-g/hw-g series digital platform scale
Hide thumbs Also See for HW-10KGL:
Table of Contents

Advertisement

14. Calibration (Adjusting the Scale)

The scale is an instrument which weighs the "weight" and displays its "mass".
Calibration is the adjustment function so that the scale can weigh correctly.
Three steps of calibration are available
Gravity Acceleration Correction ... The function to correct the scale's local gravity
Calibration of the Zero Point ......... The function to adjust the zero point, so that
Span Calibration.............................. The function to adjust the span with a
Caution
Check the accuracy of weighing periodically. Calibrate the scale, if it has
been moved to another location or the environment has changed.
Gravity acceleration correction is not required, when the scale is calibrated
with the calibration mass at the place where the scale is used.
HV-G/HW-G Series
acceleration, so that the scale functions
correctly when the calibrated scale has been
moved to a distant place.
Refer to the "Gravity Acceleration Table" on
the next page.
the zero point mark is displayed when there is
nothing on the pan.
Comment
The zero point is the fundamental starting point
to
weigh
performance of scale.
calibrated mass, so that the scale can
accurately weigh anything within the weighing
capacity.
Comment
Span means the range of weighing capacity.
Use a calibration mass heavier than two thirds
of the weighing capacity.
Page 47
anything
and
influences
14. Calibration (Adjusting Scale)
the