Percentage Mode; Storing A 100% Mass - A&D HW-10KGL Instruction Manual

Hv-g/hw-g series digital platform scale
Hide thumbs Also See for HW-10KGL:
Table of Contents

Advertisement

9. Percentage Mode

The percentage mode is the function to display a mass value in the unit of "%".
To use this function , store a 100% mass value in advance.
The 100% mass value is stored in the scale even if the power is removed.
9.1.
Step 1 Press the
Step 2 Press the
100% mass.
Step 3 With nothing on the pan, press the
display zero.
Step 4 Place the 100% mass on the pan gently.
Wait for the stability mark to be displayed. Press the
key to store. The percentage is displayed.
ENTER
Caution
When the sample is too light and it is not possible
to calculate a 100% mass, the scale displays
and returns to step 3.
Pressing the
displays the next unit.
Step 5 Remove the mass from the pan.
HV-G/HW-G Series

Storing a 100% Mass

key to display the unit %.
MODE
key to enter the mode that stores a
SET
key, after
ENTER
key to
ZERO
lo
lo
is displayed,
Page 31
Percentage mode
STABLE
Pan
STABLE
Zeroing value
100% mass
STABLE
.
Note
The pan shape depends
on the scale model.
9. Percentage Mode

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents