Basic Measurement Sequence and Example

Basic Measurement Sequence

There are five basic steps when you are making a measurement.
1. Connect the device under test and any required test equipment.
Damage may result to the device under test if it is sensitive to analyzer's
Caution
default output power level. lb avoid damaging a sensitive DUT, perform step 2
before step 1.
2. Choose the measurement parameters.
3. Perform and apply the appropriate error-correction.
4. Measure the device under test.
5. Output the measurement results.

Basic Measurement Example

This example procedure shows you how to measure the transmission response of a bandpass

Step 1. Connect the device under test and any required test equipment.

1. Make the connections as shown in Figure 2-l.

Step 2. Choose the measurement parameters.

2. Press-.
lb set preset to "Factory Preset," press:

Setting the Frequency Range.

3. lb set the center frequency to 10.24 GHz, press:
Figure 2-1. Basic Measurement Setup
Making Measurements 2-3