Verification Tests For Thermocouple Simulation - Fluke 5502E Getting Started

Multi-product calibrator
Table of Contents

Advertisement

5502E
Getting Started
Range
3.299 mF
3.299 mF
10.999 mF
10.999 mF
32.999 mF
32.999 mF
110.00 mF
110.00 mF
TC Type
10 μV/°C
TC Type
10 μV/°C
36
Table 9. Verification Tests for Capacitance (cont.)
Test Frequency
Output
540 μA dc
2.000 mF
800 μA dc
3.000 mF
900 μA dc
3.300 mF
10.900 mF
2.7 mA dc
20.000 mF
5.4 mA dc
30.000 mF
8.0 mA dc
33.00 mF
9.0 mA dc
110.00 mF
27.0 mA dc
Table 10. Verification Tests for Thermocouple Simulation
Output, °C
0.00 °C (0.0000 mV)
100.00 °C (1.0000 mV)
-100.00 °C (-1.0000 mV)
1000.00 °C (10.0000 mV)
-1000.00 °C (10.0000 mV)
10000.00 °C (100.0000 mV)
-10000.00 °C (-100.0000 mV) -100.0070
Table 11. Verification Tests for Thermocouple Measurement
Input, mV
0.00 °C (0,0000 mV)
10000.00 °C (100.0000 mV)
-10000.00 °C (-100.0000 mV) -10000.70
30000.00 °C (300.0000 mV)
-30000.00 °C (-300.0000 mV) -30001.50
Lower Limit
or Current
1.9902 mF
2.9868 mF
3.2788 mF
10.8529 mF
19.8300 mF
29.7600 mF
32.570 mF
108.800 mF
Lower Limit, mV
-0.0030
0.99696
-1.00304
9.99660
-10.0034
99.9930
Lower Limit, °C
-0.30
9999.30
29998.50
Upper Limit
2.0098 mF
3.0132 mF
3.3212 mF
10.9471 mF
20.1700 mF
30.2400 mF
33.430 mF
111.200 mF
Upper Limit, mV
0.0030
1.00304
-0.99696
10.00340
-9.9966
100.0070
-99.9930
Upper Limit, °C
-0.30
10000.70
-9999.30
30001.50
-29998.50