Ac Power Operation; Dc Power Operation; Setup - Fluke 9102S User Manual

Dry-well calibrator
Hide thumbs Also See for 9102S:
Table of Contents

Advertisement

AC Power Operation

Plug the mains power cable of the Product into a mains outlet of the proper voltage, frequency, and
current capability. Refer to
the switch on the rear panel. The dry-well turns on and heats to the previously-programmed
temperature set-point. The front-panel display indicates the actual dry-well temperature.

DC Power Operation

The Product has a dc power option. The dc option requires a power source that delivers 12 V dc at 3 A.
A fire may occur if a short circuit occurs along the input cord and no protective devices
are on the dc input source. For short-circuit protection using a battery, a fuse is
required at the battery terminals.
The dc power socket is on the rear panel of the Product near the ac power jack. The Product accepts a
7/32 inch diameter, two-conductor dc power plug such as Switchcraft® PN 760. Observe the correct
polarity as shown in Figure 1. The outside conductor is positive and the inside is negative. The ac
power switch on the rear panel does not switch the dc power.

Setup

Place the Product on a flat surface with at least 6 inches of free space around the Product. Always
leave enough clearance in front of the Product to allow for safe and easy insertion and removal of
probes. The prop stand may be swung down to raise the front of the Product from a horizontal position.
Plug the power cord into a grounded mains outlet. Observe that the nominal voltage corresponds to
that indicated on the Product.
Turn on the power to the Product with the power switch. The fan begins to quietly blow air through the
Product and the controller display illuminates after 3 seconds. After a brief self-test the Product begins
normal operation. If the unit fails to operate, check the power connection.
The display shows the well temperature and the well heater brings the temperature of the well to the
set-point temperature.
After Product use, allow the well to cool by setting the temperature to 25 °C and then wait
1/2 hour before you turn off the Product.
5
Unpacking
and Specifications, for power details. Turn on the dry-well using
+
Figure 1. 12 V dc Power Source Polarity
W
Warning
Dry-well Calibrator
Quick Start