Pace PRC 2000 Operation & Maintenance Manual page 53

Hide thumbs Also See for PRC 2000:
Table of Contents

Advertisement

6. Press any key except the TIP SET Key as the first
key of the Password.
7. Press any key as the second key of the Password.
8. Press any key as the third (and last) key of the
Password.
The Calibration Mode has been automatically entered and the
Digital Readout will now display "CAL". If no calibration
parameters need to be changed, exit Calibration Mode and go
to regular operation by pressing the TIP OFFSET Key.
Otherwise, perform calibration now (refer to the
CALIBRATION/SYSTEM SETUP section). When finished
doing calibration, press the TIP OFFSET Key to leave "CAL"
Mode and enter Temperature Display Mode (normal operation).
The Password is not permanently stored until the Calibration Mode ("CAL")
is exited normally by pressing the TIP OFFSET Key. If the POWER Switch
is turned OFF ("0") while still in the Calibration Mode, the Password will not
be saved.
Figure 54. First Key Not Tip Set Key
Figure 55. Cal Mode
NOTE
53