21. Press the TIP SET Key. Notice that the system has
retained the stored Set Tip Temperature in memory.
Immediately press the TIP SET Key once again to
exit TIP SET Mode.
22. In order to prevent a handpiece/tip combination from
inadvertently operating at an incorrect Tip
Temperature, a safety feature incorporated within the
system will prevent retention of a stored TIP
OFFSET CONSTANT if a handpiece is disconnected.
The TIP OFFSET CONSTANT will return to the
default value of "3" for °C ("6" for °F). Disconnect the
handpiece connected to the Current Channel.
Reconnect the handpiece to the same channel.
Notice that the Current Channel changes to the next
Active Channel.
23. Press the CH SELECT Key, as necessary, to
change the Current Channel to the channel
disconnected in step #22.
Figure 43. Store Set Tip Temperature
Figure 44. Handpiece Connection , TMC
Figure 45. Changing Current Channel