5. Wet the new clot filter with water and place it into the filter adapter fitting.
Orientation of the filter is not important.
6. Reconnect the filter adapter fitting to the selector valve. Finger tighten.
IMPORTANT
Be very careful not to misthread the fitting. Misthreading could strip the
valve which will prevent proper operation.
7. Run a sample and visually check for air bubbles in the sample line between
the selector valve and the UFC block.
8. Check aspiration rate.
9. If air bubbles appear, check that the filter adapter fitting is screwed on
correctly and that it is finger-tight, then repeat step 7. If air bubbles still
appear, disconnect the filter adapter fitting and check for dust or debris in the
fitting and in the port.
If air bubbles still persist, call Siemens technical support.

Replacing the Diaphragm Pumps

Time: Replacement - 5 minutes
Materials required:
1
2
3
4
5
Analyzer mode: Off
1. Remove the reagent bottles from the side where you are replacing the
diaphragm pump.
2. Pull the pump forward until it is released from its keyhole slots.
3. Remove the three tubes, then attach these tubes to the new pump.
4. Insert the raised studs on the top of the pump into the keyholes and push the
pump back until it is secure.
5. Replace the reagent bottles.
6. Checkout
5-26
Checkout - 15 minutes
DP1 - PEROX SHEATH
DP2 - SHEATH /RINSE
DP3 - SHEATH /RINSE
DP4 – WASH
PN 067-B294-04
DP5 - SHEATH /RINSE
PN 067-B294-01
For all replaced pumps - prime reagent lines and verify that there are no leaks.
REMOVE BOTTLES TO ACCESS DIAPHRAGM PUMPS
Troubleshooting the Analyzer