System Method Information: Storage And Stability; System Method Information: Sample Handling; System Method Information: Materials Required But Not Provided; System Method Information: Procedure - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

System Method Information: Storage and Stability

Reagent and calibrator storage and expiration dates are printed on the product
label. When stored unopened these products are stable through the last day of the
month stated in the expiration date, unless stated otherwise.

System Method Information: Sample Handling

Whole blood specimens whose parameter values exceed the analytical range may
be diluted with autologous plasma or isotonic saline and reassayed unless stated
otherwise in the method-specific information.
System Method Information: Materials Required but not
Provided
This topic lists the additional materials, other than the reagents, that are required
to perform the methods.

System Method Information: Procedure

For general operating instructions, refer to the Daily Routine.

System Method Information: Quality Control

This section specifies recommended quality control material and quality control
frequency for the individual method. In general, Siemens recommends that
ADVIA 2120/2120i Hematology Systems be monitored using the ADVIA
TESTpoint Hematology Controls (Low, Normal, and High) and ADVIA
TESTpoint Reticulocyte Control (Low and High). Please refer to page 5 for
product descriptions.
Control materials should be assayed at the beginning of each shift or at some
other interval chosen by the laboratory, after a reagent lot number change, and
after replacement of any part or component of the analytical module that may
affect analytical performance.
The laboratory must evaluate all control results before reporting patient results. If
control results fail to meet the laboratory's established criteria for acceptability,
all patient test results obtained in the unacceptable test run must be evaluated to
determine if patient test results were adversely affected. The laboratory should
take and document appropriate corrective actions, which may include
recalibration and reassaying of patient samples, before reporting patient results.

System Method Information: Gain Adjustment

The gain adjustment procedure is used to adjust the amplification of signals in a
channel to properly position the cell signatures within a cytogram.
Materials Required for Adjusting Gains
Whole Blood
ADVIA 120 OPTIPOINT
Regulatory Information
9-15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia 2120i

Table of Contents